Home About Us Services ↳ Canada PR Visa (Permanent Residency) ↳ Work Permit Canada ↳ LMIA — Labour Market Impact Assessment ↳ Spouse & Family Sponsorship Visa ↳ Student Visa Canada ↳ Visitor Visa ↳ Business Visa Provinces ↳ 🏙️ Ontario ↳ 🏔️ British Columbia ↳ 🌾 Alberta ↳ 🌻 Saskatchewan ↳ 🌊 Manitoba ↳ ⚓ Nova Scotia ↳ 🍁 New Brunswick ↳ 🦞 Prince Edward Island ↳ 🐟 Newfoundland & Labrador ↳ 🌊 Atlantic Immigration Program Healthcare Blog FAQ Careers Canada Contact

Reference method for measuring the surface tension of chromium electroplating: chapter 9


9. Cleaning Stalagmometers

  • 9.1 Daily cleaning:

    After completing the surface tension measurement of the solutions (at any time of the day that surface tension is to be measured), the stalagmometer must be properly rinsed with distilled water until no colour remains. Note: This may require several rinses.

    After rinsing, place the stalagmometer horizontally in a tray of fresh distilled water and allow the stalagmometer to soak in the distilled water until next use.
  • 9.2 If, after rinsing, beading of liquid occurs inside the stalagmometer, this is an indication that the stalagmometer is dirty. Beading can occur due to insufficient or improper rinsing of the stalagmometer; if the solution has been allowed to dry within the stalagmometer; or if oil or another material has coated the inside. If this situation occurs, more aggressive measures using nitric or sulphuric acid must be used to clean the stalagmometer. Use the following procedure to clean the stalagmometer:
    • 9.2.1 Place a piece of 0.5 cm") (inside diameter) tubing on the top of the stalagmometer. Place a tubing clamp on the tubing between one and two centimetres from the stalagmometer.
    • 9.2.2 Use the pipette bulb to draw up 50% nitric or 25% sulphuric acid (prepared as per subsection 3.2) to above the upper line marking on the stalagmometer. CAUTION: These acids are corrosive and produce a lot of heat when mixing with water. Wear personal protective equipment at all times: goggles, acid-resistant gloves, chemical apron and lab coat.
    • 9.2.3 Close the clamp and ensure that no liquid is released. Allow the acid to remain in the stalagmometer for one to two hours. More time may be required for very dirty equipment.
    • 9.2.4 Drain the acid from the stalagmometer and rinse the stalagmometer at least three times with distilled water.
    • 9.2.5 If oil is suspected to cause the beading, then, after soaking the stalagmometer in acid, rinse once with distilled water, then rinse once with full strength acetone or methanol, and finally rinse at least three times with distilled water.

Page details

2017-09-10

Quick Enquiry

We usually reply within a few hours
By submitting you agree to be contacted about your enquiry.
Call us Chat on WhatsApp
M

Migova AI Assistant

Online now
Hi 👋 I'm the Migova AI assistant, powered by OpenAI. Ask me about PR, study visas, work permits, LMIA, family sponsorship, provinces, or healthcare immigration to Canada.
Canada PR
Study Visa
LMIA / Work Permit